CASE LETTER
Year : 2016  |  Volume : 33  |  Issue : 1  |  Page : 103-105  

“Undulation on ventilator wave” may indicate serious lung pathology


Department of Anesthesia and Critical Care, Nazareth Hospital, Shillong, Meghalaya, India

Date of Web Publication4-Jan-2016

Correspondence Address:
Saurabh Kumar Das
Department of Anesthesia and Critical Care, Nazareth Hospital, Shillong, Meghalaya
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-2113.173081

Rights and Permissions

How to cite this article:
Das SK, Chopoo NS. “Undulation on ventilator wave” may indicate serious lung pathology. Lung India 2016;33:103-5

How to cite this URL:
Das SK, Chopoo NS. “Undulation on ventilator wave” may indicate serious lung pathology. Lung India [serial online] 2016 [cited 2020 Feb 22];33:103-5. Available from: http://www.lungindia.com/text.asp?2016/33/1/103/173081

Sir,

Real-time examination of ventilator waves and loops helps us understand respiratory mechanics, patient ventilator interactions, and other pathophysiological conditions.[1] Therefore, ventilator waves and loops have become an integral part of intensive care and operation theater monitoring. We would like to describe an interesting characteristic of pressure and flow time wave of three patients, which turned out to be due to a serious, life threatening lung pathology.

A 58-year-old male who was suffering from multidrug-resistant tuberculosis with pneumothorax was admitted to the intensive care unit (ICU) with severe respiratory difficulty. There was a chest tube with water-seal drainage on the right side, and movement of air-fluid level could be appreciated. He was intubated, and ventilation (Ventilator: eXtend, Air liquid, Paris, France) was started with synchronized intermittent mandatory ventilation plus pressure support (SIMV + PS) at 100% fraction of inspired oxygen (FiO2) and positive end-expiratory pressure (PEEP) of 3 cm H2O. As the patient required high FiO2, PEEP was increased gradually. At that moment, we observed an oscillatory wave in the expiratory phase of pressure and flow time wave resembling the undulation that is usually noticed due to secretion in the ventilator circuit [Figure 1] and [Figure 2]. The endotracheal tube was suctioned, but it was dry and the undulation persisted. When we decreased PEEP to 3 cm H2O, the wave disappeared. But it reappeared again as PEEP was increased. The same phenomenon was observed several times. Portable chest x-ray showed a right-sided pneumothorax with a kinked chest tube in situ. The pneumothorax was decompressed with a new 24-Fr chest tube. As soon as the pneumothorax was decompressed, the undulation disappeared. The patient's condition gradually improved and the trachea was extubated the next day.
Figure 1: Absence of oscillatory waves on expiratory phase of pressure and flow-time wave at PEEP 3

Click here to view
Figure 2: Oscillatory waves on expiratory phase of pressure and flow-time wave at PEEP 4

Click here to view


The second case involved an 11-year-old girl who was admitted because of severe community-acquired pneumonia with left-sided pneumothorax. The pneumothorax was decompressed and the patient was placed on mechanical ventilation (ventilator :Maquet Servo I, Andheri, Mumbai). On the third day of ventilation, oscillatory waves on pressure and flow-time waves were observed. But in this case, the amplitudes of the oscillatory waves were small. We could appreciate the waves only when PEEP was higher than 7 cm H2O, and waves were absent when there was no PEEP [Figure 3]. Chest x-ray showed right-sided pneumothorax. Undulation disappeared as the pneumothorax was decompressed by another chest tube on the right side. Unfortunately, after a few days the child died due to severe sepsis and multiple organ failure.
Figure 3: Small oscillatory waves on expiratory phase of pressure and flow-time wave at PEEP 7

Click here to view


The third case was of a 27-year-old patient who had sustained head and multiple chest injuries following a road traffic accident. The patient had a left-sided pneumothorax, which was drained by a 28-Fr chest tube and water-sealed drainage. Two hours after inserting the tube, the patient developed subcutaneous emphysema along with undulation on the expiratory phase of pressure and flow-time wave of the ventilator, which varied with the amount of PEEP applied. Chest x-ray showed pneumothorax along with a kinked chest tube and the most proximal opening outside the thoracic cavity. Undulation disappeared as the chest tube was repositioned.

Various intrinsic and extrinsic factors such as ineffective respiratory effort, terminal asynchrony, secretions, and electromagnetic interference may influence ventilator display.[2],[3] This sawtooth pattern or oscillation in ventilator waves and loops is usually considered as an indicator of secretion in airways.[1],[4] During exhalation, the presence of secretion in the airway can generate cyclic increments and decrements in flow, secondary to transient obstruction of the airway lumen or the fluctuating movement of its content.[5] Extensive literature search did not reveal any association between this typical undulation and incomplete decompression of pneumothorax. Though we could not offer any definite explanation for this typical finding in these cases, we observed three interesting and common characteristics. First, endotracheal tube suction did not reveal any secretion.

Second, undulation appeared as PEEP was increased. Third, complete decompression of the pneumothorax abolished these waves permanently. However, we cannot attribute this finding to incomplete decompression of pneumothorax only on the basis of these three cases. Further observations or studies are warranted to validate this finding.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Corregr E, Murias G, Chacon E, Estruqa A, Sales B, Lopez-Aquilar J, et al. Interpretation of ventilator curves in patients with acute respiratory failure. Med Intensiva 2012;36:294-306.  Back to cited text no. 1
    
2.
Gonzalvo R, Martí-Sistac O, Blanch L, Lόpez-Aquilar J. Bench-to-bedside review: Brain-lung interaction in the critically ill--a pending issue revisited. Crit Care 2007;11:216.  Back to cited text no. 2
    
3.
Lapinsky SE, Easty AC. Electromagnetic interference in critical care. J Crit Care 2006;21:267-70.  Back to cited text no. 3
    
4.
Sokhal N, Goyal K, Chowdhury T, Rath GP. See-saw pattern in ventilator graphic: Is there any story behind? Niger Med J 2014;55:359-61.  Back to cited text no. 4
[PUBMED]  Medknow Journal  
5.
Koutsoukou A, Beskos B, Sotiropoulou C, Koulouris NG, Roussos C, Milic-Emili J. Effects of positive end-expiratory pressure on gas exchange and expiratory flow limitation in adult respiratory distress syndrome. Crit Care Med 2002;30:1941-9.  Back to cited text no. 5
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]



 

Top
  
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed883    
    Printed4    
    Emailed0    
    PDF Downloaded135    
    Comments [Add]    

Recommend this journal